DESIGN OF MANIFOLD WITH PRESSURE CONTROLLER FOR AUTOMATIC EXCHANGE OF OXYGEN GAS CYLINDERS IN HOSPITAL

Design of Manifold with Pressure Controller for Automatic Exchange of Oxygen Gas Cylinders in Hospital

Design of Manifold with Pressure Controller for Automatic Exchange of Oxygen Gas Cylinders in Hospital

Blog Article

The regulation and supply of oxygen as one of the medical gases in the hospital is important to ensure the availability of these gases for the survival of patients.The regulation of oxygen gas in hospitals usually uses a piping system with manifolds.The manifold will old taylor whiskey 1933 price monitor the oxygen gas pressure on each tube.Manifold systems that are widely used in general can only monitor pressure but cannot perform an automatic exchange on gas cylinders if the pressure is under the permissible conditions.

The manifold system design developed is equipped with pressure monitoring for automatic exchange of oxygen gas cylinders using pressure sensors and microprocessors.The test results of the system using regulator and barometer comparisons showed the percentage value c6097b1028 of sensor pressure accuracy of 96.92 percent and 97.16 percent.

At pressure below the limit of 285 KPa manifold can perform the exchange of active gas cylinders automatically.These results show the manifold design built can work quite well.